Title |
Estimation of Systolic Blood Pressure using PTTL |
Authors |
길세기(Kil, Se-Kee) ; 권장우(Kwan, Jang-Woo) ; 윤광섭(Yoon, Kwang-Sub) ; 이상민(Lee, Sang-Min) |
Keywords |
PTTL ; Blood Pressure ; Estimation ; Regression Method ; Body Parameters |
Abstract |
The desirable method to diagnose abnormal blood pressure is to measure and manage blood pressure continuously and regularly. However, the sphygmomanometers that are based on a cuff have faults in that they can not measure the blood pressure continuously and they cause an unpleasant feeling. Therefore, it is essential to develop a new measuring method that causes no pain and that can obtain blood pressure continuously without any unpleasant feeling. Thus, we propose here a regression method to estimate the systolic blood pressure by using the PTTL(pulse transit time on leg) with some body parameters which are chosen from the relational analysis with systolic blood pressure. The data we use to make the regression model were obtained in triplicate from each of 50 males who were from 18 to 35 years. And we made estimation experiments of blood pressure on 10 males who did not take part in the making the regression model. According to the results, the proposed method showed a mean error of 4.00 mmHg and the standard variance was 2.45 mmHg. When we comparing the results of the proposed method with the rule of American National Standards Institute of the Association of the Advancement of Medical Instruments(ANSI/AAMI), the results satisfied the rule of a mean error less than 5 mmHg and a standard variance less than 8 mmHg. Therefore we were able to validate the usefulness of the proposed method. |