Title |
Development and Verification of a Hand-held Typed Monitoring System for Cardiac Output Measurement using Electrical Impedance Technique |
Authors |
서광석(Seo, Kwang-Seok) ; 심명헌(Sim, Myeong-Heon) ; 김민용(Kim, Min-Yong) ; 윤찬솔(Yoon, Chan-Sol) ; 정주홍(Chung, Joo-Hong) ; 박성빈(Park, Sung-Bin) ; 윤형로(Yoon, Hyung-Ro) |
DOI |
https://doi.org/10.5370/KIEE.2012.61.12.1932 |
Keywords |
Cardiac output ; Stroke volume ; Impedance cardiogram ; Hand-held type electrode ; Physioflow |
Abstract |
The impedance cardiogram has been proposed as a non-invasive, continuous, operator independent, and cost-effective method for cardiac output monitoring. However, it can not be completely considered as non-restrictive method because of attached spot and band type electrodes. Therefore, we developed a improved convenient hand-held typed measurement system for cardiac output by electrical impedance technique. 80 subjects from Yonsei University and the surrounding areas, participated. All subjects measured stroke volume and cardiac output through Physioflow and developed system. To verify the developed system, statistical methods such as correlation, Wilcoxon signed ranks test, and the Bland-Altman analysis were used. The proposed system showed significant correlation in both male and female stroke volume(r=0.715, r=0.704) and cardiac output(r=0.826; r=0.804). From these results, it can be concluded that stroke volume and cardiac output could be improved convenient measurement using the both hands without the help of a specialist. |