Title |
Study of the Wearable Electrocardiogram Measuring System using Capacitive-coupled Electrode |
Authors |
이재호(Lee, Jae-Ho) ; 이영재(Lee, Young-Jae) ; 이강휘(Lee, Kang-Hwi) ; 강승진(Kang, Seng-Jin) ; 김경남(Kim, Kyeung-Nam) ; 박희정(Park, Hee-Jung) ; 이정환(Lee, Jeong-Whan) |
DOI |
https://doi.org/10.5370/KIEE.2014.63.10.1448 |
Keywords |
Capacitive-coupled electrode ; ECG ; Smoothing ; Savitzky-Golay filter |
Abstract |
In this study, a new type of electrode device is implemented to measure the capacitance energy and interpret it as the ECG (Electrocardiogram) data. The main idea of this new electrode system is to estimate the capacitance on the skin by assembling a capacitive-coupled circuits and translate into the ECG signal. To measure the coupling energy and estimate the aquired data in terms of heart activity, the capacitive-coupled electrode is garmented with fabrics in the form of a chest band or a vest jacket. To compare the ECG data from the capacitive-coupled electrode with the conventional electrode(Ag-AgCl) system, the corelation coefficient between two signals is computed as 0.9517. Thus, we can conclude the fact that capacitive-coupled electrode system can measure a person's heart activity without any contact to his or her skin and can the interpreted as the ECG data. |